12/02/2016

JM 07:53 AM Support: RE: Acquiring images from MityCam CIS1910 with exposures grater than 1 sec
Anton,
The dark calibration would have been cleared when you reflashed the SD card. For best performance, you should set the clock to your desired rate, then run the calibration command with the sensor darkened. The calibration values...
Jeffrey Myers
AB 05:19 AM Support: RE: Acquiring images from MityCam CIS1910 with exposures grater than 1 sec
Update. It return non zero images only for exposures bigger than 6 sec.
With exposures in interval 6.01 - 6.1 sec images are partially zero (lower part of image). More exposure - less zero valued part.
The same issue on 80MHz also.
Anton Bakulev
AB 03:07 AM Support: RE: Acquiring images from MityCam CIS1910 with exposures grater than 1 sec
Jeffrey,
Thank you for your reply. I've upgraded a firmware successfully. Now it reports the following:
Anton Bakulev
AB 03:01 AM Support: RE: There is no eth0 device in 40-000065-4 release firmware
Jeffrey,
Thank you. Yes, this instructions fixes the problem. Now my MityCam works well by eth0.
Anton Bakulev

12/01/2016

JM 07:56 AM Support: RE: There is no eth0 device in 40-000065-4 release firmware
Anton,
I sent you an email with instructions on reconfiguring the camera.
Jeffrey Myers
AB 04:44 AM Support: There is no eth0 device in 40-000065-4 release firmware
Hello,
I've installed a 40-000065-4 release firmware from [[https://support.criticallink.com/redmine/projects/mityvision/wiki/MityCAM-B1910_Firmware]] (by the way, file name is MityCAM_B1910-40-000064-4_soc_camera.img.zip, not a MityC...
Anton Bakulev

11/29/2016

AB 07:40 AM Support: RE: Acquiring images from MityCam CIS1910 with exposures grater than 1 sec
Thank you.
Yes, it's fully enclosed camera that was re-flashed several months ago in this post? [[https://support.criticallink.com/redmine/boards/21/topics/4794]]
Ok. I'll update a firmware.
Anton Bakulev
JM 07:29 AM Support: RE: Acquiring images from MityCam CIS1910 with exposures grater than 1 sec
Anton Bakulev wrote:
> I've tried to set clock rate to 40MHz but images acquired after that is zero valued. Did you tried this approach on your site?
The MityViewer uses this approach to adjust the clock when changing between rolling...
Jeffrey Myers
AB 05:42 AM Support: RE: Acquiring images from MityCam CIS1910 with exposures grater than 1 sec
I've tried to set clock rate to 40MHz but images acquired after that is zero valued. Did you tried this approach on your site?
Also I've found that defailt clock rate is 283Mhz. Such a value was received after hard reboot of MityCam.
A...